Evaluation of three fully automated immunoassay systems for detection of IgA anti-beta 2-glycoprotein I antibodies. [artículo]
Por: Pérez Méndez, Dolores [Inmunología] | Martinez-Flores, Jose A [Instituto de Investigación i+12] | Serrano, Manuel [Instituto de Investigación i+12] | Lora Pablos, David [Instituto Investigación I+12] | Paz Artal, Estela [Inmunología] | Morales, José M [Instituto de Investigación imas12] | Serrano Hernández, Antonio [Inmunología].
Colaborador(es): Instituto de Investigación imas12 | Servicio de Inmunología.
Tipo de material: ArtículoEditor: International journal of laboratory hematology, 2016Descripción: 38(5):560-8.Recursos en línea: Solicitar documento Resumen: Introduction: In recent years, we have been witnessing increased clinical interest in the determination of IgA anti-beta 2-glycoprotein I (aB2GPI) antibodies as well as increased demand for this test. Some ELISA-based diagnostic systems for IgA aB2GPI antibodies detection are suboptimal to detect it. The aim of our study was to determine whether the diagnostic yield of modern detection systems based on automatic platforms to measure IgA aB2GPI is equivalent to that of the well-optimized ELISA-based assays. Methods: In total, 130 patients were analyzed for IgA aB2GPI by three fully automated immunoassays using an ELISA-based assay as reference. The three systems were also analyzed for IgG aB2GPI with 58 patients. Results: System 1 was able to detect IgA aB2GPI with good sensitivity and kappa index (99% and 0.72, respectively). The other two systems had also poor sensitivity (20% and 15%) and kappa index (0.10 and 0.07), respectively. On the other hand, kappa index for IgG aB2GPI was >0.89 in the three systems. Conclusion: Some analytical methods to detect IgA aB2GPI are suboptimal as well as some ELISA-based diagnostic systems. It is important that the scientific community work to standardize analytical methods to determine IgA aB2GPI antibodies.Tipo de ítem | Ubicación actual | Signatura | Estado | Fecha de vencimiento |
---|---|---|---|---|
Artículo | PC17814 (Navegar estantería) | Disponible |
Navegando Hospital Universitario 12 de Octubre Estantes Cerrar el navegador de estanterías
Formato Vancouver:
Pérez D, Martínez Flores JA, Serrano M, Lora D, Paz Artal E, Morales JM et al. Evaluation of three fully automated immunoassay systems for detection of IgA anti-beta 2-glycoprotein I antibodies. Int J Lab Hematol. 2016 Oct;38(5):560-8.
PMID: 27435259
Contiene 39 referencias
Introduction: In recent years, we have been witnessing increased clinical interest in the determination of IgA anti-beta 2-glycoprotein I (aB2GPI) antibodies as well as increased demand for this test. Some ELISA-based diagnostic systems for IgA aB2GPI antibodies detection are suboptimal to detect it. The aim of our study was to determine whether the diagnostic yield of modern detection systems based on automatic platforms to measure IgA aB2GPI is equivalent to that of the well-optimized ELISA-based assays.
Methods: In total, 130 patients were analyzed for IgA aB2GPI by three fully automated immunoassays using an ELISA-based assay as reference. The three systems were also analyzed for IgG aB2GPI with 58 patients.
Results: System 1 was able to detect IgA aB2GPI with good sensitivity and kappa index (99% and 0.72, respectively). The other two systems had also poor sensitivity (20% and 15%) and kappa index (0.10 and 0.07), respectively. On the other hand, kappa index for IgG aB2GPI was >0.89 in the three systems.
Conclusion: Some analytical methods to detect IgA aB2GPI are suboptimal as well as some ELISA-based diagnostic systems. It is important that the scientific community work to standardize analytical methods to determine IgA aB2GPI antibodies.
No hay comentarios para este ejemplar.