A European multicenter study on the analytical performance of the VERIS HBV assay. [artículo]
Por: Delgado Vázquez, Rafael [Microbiología y Parasitología].
Colaborador(es): Servicio de Microbiología y Parasitología | Instituto de Investigación imas12.
Tipo de material: ArtículoEditor: Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology, 2015Descripción: 99-100:50-56.Recursos en línea: Solicitar documento Resumen: Background: Hepatitis B viral load monitoring is an essential part of managing patients with chronic Hepatits B infection. Beckman Coulter has developed the VERIS HBV Assay for use on the fully automated Beckman Coulter DxN VERIS Molecular Diagnostics System.1 OBJECTIVES: To evaluate the analytical performance of the VERIS HBV Assay at multiple European virology laboratories. Study design: Precision, analytical sensitivity, negative sample performance, linearity and performance with major HBV genotypes/subtypes for the VERIS HBV Assay was evaluated. Results: Precision showed an SD of 0.15 log10 IU/mL or less for each level tested. Analytical sensitivity determined by probit analysis was between 6.8-8.0 IU/mL. Clinical specificity on 90 unique patient samples was 100.0%. Performance with 754 negative samples demonstrated 100.0% not detected results, and a carryover study showed no cross contamination. Linearity using clinical samples was shown from 1.23-8.23 log10 IU/mL and the assay detected and showed linearity with major HBV genotypes/subtypes. Conclusions: The VERIS HBV Assay demonstrated comparable analytical performance to other currently marketed assays for HBV DNA monitoring.Tipo de ítem | Ubicación actual | Signatura | Estado | Fecha de vencimiento |
---|---|---|---|---|
Artículo | PC16728 (Navegar estantería) | Disponible |
Navegando Hospital Universitario 12 de Octubre Estantes Cerrar el navegador de estanterías
Formato Vancouver:
Braun P, Delgado R, Drago M, Fanti D, Fleury H, Izopet J et al. A European multicenter study on the analytical performance of the VERIS HBV assay. J Clin Virol. 2018 Feb-Mar;99-100:50-56.
PMID: 29328964
Contiene 33 referencias
Background: Hepatitis B viral load monitoring is an essential part of managing patients with chronic Hepatits B infection. Beckman Coulter has developed the VERIS HBV Assay for use on the fully automated Beckman Coulter DxN VERIS Molecular Diagnostics System.1 OBJECTIVES: To evaluate the analytical performance of the VERIS HBV Assay at multiple European virology laboratories.
Study design: Precision, analytical sensitivity, negative sample performance, linearity and performance with major HBV genotypes/subtypes for the VERIS HBV Assay was evaluated.
Results: Precision showed an SD of 0.15 log10 IU/mL or less for each level tested. Analytical sensitivity determined by probit analysis was between 6.8-8.0 IU/mL. Clinical specificity on 90 unique patient samples was 100.0%. Performance with 754 negative samples demonstrated 100.0% not detected results, and a carryover study showed no cross contamination. Linearity using clinical samples was shown from 1.23-8.23 log10 IU/mL and the assay detected and showed linearity with major HBV genotypes/subtypes.
Conclusions: The VERIS HBV Assay demonstrated comparable analytical performance to other currently marketed assays for HBV DNA monitoring.